VEGA VEGAFLEX 82 4 … 20 mA_HART four-wire User Manual
Page 22
22
5 Connecting to power supply
VEGAFLEX 82 • 4 … 20 mA/HART - four-wire
41830-EN-130314
Fig. 11: Connection steps 5 and 6
6. Insert the wire ends into the terminals according to the wiring plan
Information:
Solid cores as well as flexible cores with wire end sleeves are
inserted directly into the terminal openings. In case of flexible cores
without end sleeves, press the terminal head with a small screwdriver;
the terminal opening is freed. When the screwdriver is released, the
terminal closes again.
You can find further information on the max. wire cross-section under
"Technical data/Electromechanical data"
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connect the screen to the internal ground terminal, connect the
outer ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
10. Place probably existing display and adjustment module back on
11. Screw the housing cover back on
The electrical connection is hence finished.