5 technical data – Xylem CG 842 User Manual

Page 10

Advertising
background image

Overview

53

1.5

Technical data

Measuring accuracy

(± 1 digit)

Given as measuring uncertainty at a confidence level of
95%. All accuracies additionally depend on the accuracies
of the measuring probes.

Storage temperature

- 25 °C ... + 65 °C

Operating temperature

0 °C ... + 55 °C

Allowable relative hu-
midity

Annual mean: < 75 %
30 days/year: 95 %
Other days: 85 %

pH
(resolution selectable)

- 2.000 ... + 16.000
- 2.00 ... + 16.00

U [mV]

- 999.9 ... + 999.9
- 1999 ... + 1999

T [°C]

- 5.0 ... + 100.0

T [°F]

+ 23.0 ... + 212.0

Ambient

temperature

Measuring ranges

and resolution

pH
(± 2 pH units around the
calibration point)

± 0.005 at + 15 °C ... + 35 °C
± 0.01

U [mV]

± 0.4 at + 15 °C ... + 35 °C
± 1

T [°C]

NTC 30: ± 0.1

PT 1000:
± 0.5 at 0 °C ... 15 °C
± 0.1 at 15 °C ... 35 °C
± 1 at 35 °C ... 55 °C

T [°F]

NTC 30: ± 0.2

PT 1000:
± 0.9 at 32 °F ... 59 °F
± 0.2 at 59 °F ... 95 °F
± 1.8 at 95 °F ... 131 °F

Advertising