Operation
76
4.4
Reset
Basic settings
The following functions are reset (initialized) to the values
they had on delivery:
Proceed as follows:
Measuring mode
pH
Asymmetry 0
mV
Slope -59.16
mV/pH
Calibration procedure
AutoCal TEC
Temperature, manual
25°C
Resolution of pH display
0,01
1 Press and hold down the <run/enter> key.
2 Press the <CAL> key.
3 Toggle between no and yes by pressing <
▲
> <
▼
>.
yes: reset parameters.
no: retain settings.
4 Confirm with <run/enter>.
The pH meter changes automatically to the pH mea-
suring mode.