1 measuring the ph value – Xylem CG 842 User Manual
Page 22
Operation
65
4.2.1 Measuring the pH value
AutoRead AR
(Drift control)
The
AutoRead
function (drift control) checks the stability of
the measurement signal. The stability has a considerable ef-
fect on the reproducibility of the measured values.
For identical measurement conditions, the following criteria
apply:
l
pH value: better than 0.02 (setting time: > 30 s)
1 Perform the preparatory activities according to sec-
2 Immerse the pH electrode into the test sample.
3 Press the <pH> key until
pH
appears in the status
display. The pH value appears on the display.
1 Call up the pH measuring mode by pressing <pH>.
2 Activate the AutoRead function by pressing <auto
read>.
The current measured value is frozen (Hold function).
3 Start the AutoRead function by pressing
<run/enter>.
AR flashes on the display until a stable measured val-
ue is reached.
4 If necessary, start the next AutoRead measurement
by pressing <run/enter>.
5 To terminate the AutoRead function: Press the <auto
read> key.