4.2.1 Measuring the pH value

AutoRead AR

(Drift control)

The

AutoRead

function (drift control) checks the stability of

the measurement signal. The stability has a considerable ef-
fect on the reproducibility of the measured values.

For identical measurement conditions, the following criteria
apply:

l

pH value: better than 0.02 (setting time: > 30 s)

1 Perform the preparatory activities according to sec-

tion 4.2.

2 Immerse the pH electrode into the test sample.

3 Press the <pH> key until

pH

appears in the status

display. The pH value appears on the display.

1 Call up the pH measuring mode by pressing <pH>.

2 Activate the AutoRead function by pressing <auto

read>.
The current measured value is frozen (Hold function).

3 Start the AutoRead function by pressing

<run/enter>.
AR flashes on the display until a stable measured val-
ue is reached.

4 If necessary, start the next AutoRead measurement

by pressing <run/enter>.

5 To terminate the AutoRead function: Press the <auto

read> key.