Operation

71

4.3.2 AutoCal DIN

Use two different DIN buffer solutions for this procedure:

Note
Steps 2 and 6 are not required if you use a temperature
probe.

DIN buffer solution

pH

A

1.68

C

4.01

D

6.87

F

9.18

1 Press the <CAL> key repeatedly until Cd1 and the

AutoCal DIN function display appear.

2 If necessary, set the temperature of the buffer solu-

tion by pressing <

> <

>.

3 Submerse the pH electrode in the first buffer solution.

4 Press the <run/enter> key.

AR flashes on the display.
The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized, Cd2 ap-
pears.