1 autocal tec – Xylem CG 842 User Manual

Page 26

Advertising
background image

Operation

69

4.3.1 AutoCal TEC

Use any two of the Technical Buffer Solutions for this proce-
dure.

Note
Steps 2 and 6 are not required if you use a temperature
probe.

1 Press the <CAL> key repeatedly until Ct1 and the

AutoCal TEC function display appear.

2 If necessary, set the temperature of the buffer solu-

tion by pressing <

> <

> .

3 Submerse the pH electrode in the first buffer solution.

4 Press the <run/enter> key.

AR flashes on the display.
The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized, Ct2 ap-
pears.

5 Thoroughly rinse the electrode with distilled water.