Symptom Possible

Cause

Remedy

Err 2

• Over-load or Under-load

condition.

• Transportation Lock is set
• Improper calibration

• Verify pan is installed and/or excess load is

removed from pan.

• Slide the Lock pin to the unlocked position.
• Perform calibration.

Err 4

• Invalid calibration mass.

• Verify correct calibration mass is being used.

Err 8

• RS232 buffer is full.

• Cycle scale Off/On.
• Check RS232 settings in scale and computer

when RS232 option is installed.

Err 9

• Internal data error.

• Contact an authorized dealer.

(Flashing)

• Battery power is low.

• Replace batteries soon

bAtt LowW

• Battery power is too low for
proper operation.

• Replace batteries

5.2 Service Information

If the troubleshooting section does not resolve or describe your problem, contact your
authorized dealer.

6. TECHNICAL DATA

6.1 Specifications

TABLE 6-1.SPECIFICATIONS

Capacity (g)

100 300 150 300 300 500 900 1500 3000 5000 6000

Readability (g)

0.001 0.001 0.01

0.01 0.1 0.1 0.1 0.1 0.1 1 0.1

Linearity (d)

±3 ±5 ±1 ±1 ±1 ±1 ±1 ±1 ±1 ±1 ±2

Repeatability (g)

0.003 0.005 0.01

0.01 0.1 0.1 0.1 0.1 0.1 1 0.1

Span Mass (g)

100 200 100 200 200 300 500 1000 2000 3000 3000

Linearity Masses (g)

50

100

200
300

100
150

200
300

200
300

300
500

500
900

1000
1500

2000
3000

3000
5000

3000
5000

Weighing Units

Gram, Newton

Tare Range

To Capacity by Subtraction

Stabilization Time (sec)

< 3

Operating Temperature Range

50° - 104°F / 10° - 40°C

Operating Humidity Range

30%-90%

Pan Size (mm)

90 dia.

120 dia.

123 x 124