4.2.1 Setup Menu

[

S.E.t.U.P.]

The Setup menu contains the following sub-menus:

Auto-Off [.

A-OFF.]:

Auto Off settings:

On Off

Setting

[On] will automatically turn the scale off after no

activity is detected for 4 minutes.


Span Calibration [

SPAN]: Uses two weight values: Zero and a value at or near the full

capacity of the scale. Span calibration should be performed
if the scale is repositioned or if the room temperature
changes significantly.


Linearity Calibration [

Lin]: Uses three weight values: Zero, a mid-range value and full

capacity. Generally this calibration is not required unless
testing shows that the linearity error exceeds the linearity
tolerance in the specification table. Accuracy of the weights
is important to maintain weighing integrity.

4.2.2 Unit Menu

[

U.n.i.t]

Each Weighing Unit may be individually set to

On or OFF.

See Table 4-1 for the symbols associated with each Unit or Mode.

4.2.3 Other Menus

Additional Menus are available if a serial interface option is installed in the scale. These
menus are explained within the user manual of each option kit.

4.3. Calibration Lock Feature

Sliding the switch fully to the lock position prevents the calibration of the scale. Using
a paper or wire seal will physically secure the switch if required.



Locking the calibration

Adhesive paper seal

Wire or plastic tie seal

Figure 4-1. Locking and Sealing the Calibration Lock Switch.