Weighing mode, Operation, 6 initial calibration – Carolina Electronic Balances User Manual

Page 7: 1 draft shield, 2 weighing mode

Advertising
background image

2.6 Initial Calibration

Power on the scale by pressing On-Zero. Allow at least one minute for the scale to
stabilize before calibrating. Press and hold the Print/Cal button until [

CAL] is

displayed. The display flashes [

-C-] while zero reading is taken. The display flashes

the required span mass. Put the indicated calibration mass on the pan and press On-
Zero.
The display flashes[

-C-]. When calibration is complete, [DONE] is displayed.

Note: Calibration is also available in the Setup menu, see section 4.2.1.

3. OPERATION

Note that one or more weighing Units of measure may initially be unavailable. They can
be enabled in the Main menu Unit

[U.n.i.t] , see section 4.2.2.

3.1 Draft Shield

The draft shield was designed with an easy to remove center piece which can be used
as a high quality sample tray. Even with the center piece removed the draft shield will
provide increased protection from rough environments while maintaining weighing
speed. To remove the center piece, rotate counter-clockwise. See Figure 3-1. With the
sample tray firmly in place the unique shape of the draft shield also allows for efficient
stacking of one scale on top of another.














Figure 3.1, Sample Tray Removal

3.2 Weighing Mode

Press Unit to advance to next available weighing Unit.
Press Zero to zero the scale and then place objects to be weighed on the pan.