Inclinometer RS232

– Current Datasheet

Version: 20130503

INCLI NO MET E R

RS2 32

– CURRENT

9 Commands in User Level

Instruction

To the

sensor

Response sensor

Description

activate temporary polling

mode

„f“

„f“

the continuous sending of

angle values is stopped,

instructions can be sent to the

sensor

activate temporary continuous

mode

„F“

„X=vxx.xxx“, CR, LF,
„Y=vxx.xxx“, CR, LF,
„X= . . .

X angle in °

Y angle in °
with „±” = „+“ or „-“,

one string contains x and y

value

read angle values

at one-timeErreur ! Source

du renvoi introuvable.

„R“

„X=vxx.xxx“, CR, LF,
„Y=vxx.xxx“, CR, LF,

X angle in °
Y angle in ° with „±” = „+“ or „-“

switch to the setup level

3) 4)

”prog”

”P“

Sensor is at setup level

show active level

“*“

“Ux“ or
”Sx“

„U“ means Sensor is at User

level
„S“ means Setup-level is

active,
with „x“ Output-Mode of

Sens

ors „U“ / „I“ / „P“ / „S“

1) In free running mode measurement data is continuously displayed. In query mode measurement and display is only once on

command.

2) After reset or new Power On after an interruption of power supply, the sensor will be in user-level again with the original setup or with

the setup changed in the setup level.

3) Only possible in query mode (=free running mode deactivated).

4)

The Input of „prog“ must take place within 20 sec.