BEI Sensors High Resolution Inclinometers User Manual

Page 12

Advertising
background image


Page 12

Inclinometer RS232

– Current Datasheet

Version: 20130503

INCLI NO MET E R

RS2 32

– CURRENT

10 Setup Level

The Setup level is active until Power On or Reset.

All settings taken in the setup level are stored in

the EEPROM and permanent available also after

Power down.

Instruction

To the sensor

Response sensor

Description

activate permanent

polling mode

1)

“f“

“f“

the continuous sending of

angle values are

permanent stopped,

instructions can send to

the sensor

activate permanent

continuous mode

1)

“F“


„X=±xx.xxx“, CR, LF,
”Y=±xx.xxx“, CR, LF,
”X= . . .

continuous sending of

X angle in °

Y angle in °
with „±” = „+“ or „-“

set rate of data

transmission for

continuous mode

2) 3) 4)

“O“

<Code transmission

rate>

“O”

<Code transmission

rate>

Echo,

Code transmission rate
or „E“ for Error, if the

code is outside defined

values

read angle values at

one-time

2)

“R“

same as at user level

read version

2)

“V“

“AGSxxx-2-Sx“, CR,

LF
”SN:xxxx-xxx“, CR,

LF
”HV:xx.x“ , CR, LF
”SV:xx.x“ , CR, LF

type of Sensor

serial number

HW Version internal

sensor

SW Version

offset adjust of the

specified axis

2) 3)

“n“

”x“ or ”y“

“n“
”OffsetX=±xx.xxx“ or
”OffsetY=±xx.xxx“

the actual angle of

specified axis is set to

zero, ±xx.xxx is the

internal offset in

degree

reset offset adjust

2) 3)

“N“

“N“

the offset adjust was

reset to the original value

Set Baud rate

2) 3) 6)

“B“

<Code Baud rate>

“B“

<Code Baud rate>

Echo, Code Baud rate or
„E“ for Error, if the code

is outside defined values

Filter deactivated

„M0“

„M0“

all filters are deactivated

Moving Average

Filter

10)

„M1“, “mm“

„M1mm“

Filter: Moving Average

Filter is activated