4-3 Indicator Reference Mark (Zero) Adjustment

This adjustment is performed at a no flow condition with the float at rest.

1. Completely stop fluid flow through the meter.
2. Loosen the screw on the scale.
3. Align the "Ref" (zero) mark on the scale with the line on the indicator.
4. Tighten the scale screw.

4-4 Modify Alarm Set-Point

1. Unscrew and remove alarm housing cover.
2. Loosen screw in reed switch holder.
3. Required alarm setting can be achieved by sliding reed switch holder

up or down on isolation well until desired alarm position is reached.

4. Tighten screw in reed switch holder securely.
5. Replace cover.