Se 56 – Burkert Type 8056 User Manual

Page 35

Advertising
background image

SE 56

35

(POS. 6.5-6) Function and the range of current output n.1-2

[Out mA1-2=X÷XX+]

The function associated to the signal current on output N.1-2. The current output N.1 is optional
and it is
mounted on the main board. There are three fields to modify for this function:

‰

Scale zero: 4 or 0mA;

‰

Full scale: 20 or 22mA

‰

Field: + = positive, - = negative, blank = both, -0+ = central zero scale

The values corresponding to the scale points are shown in the following chart:












In hardware alarm conditions “HARDW AL.” (interrupted coils, empty pipe, measure error) the current
value is configured by the function “mA v.fault” (pos. 4.6) and it is expressed as percentage of a fixed
current range, where: 0% = 0mA and 110% = 22mA.













MENU 8 - DISPLAY

(POS. 8.3) Display contrast set

[Contrast=X]

Visual display contrast set. The contrast can change according to the room temperature. The set
values are from 0 to 15. The entered value has effect only when leaving the function itself.
WARNING >10 is a dark screen and obscures writing on the display
Contrast also can be set from visualization pages pushing the key for 8 seconds or
more. In this way the contrast set that will be visualized at release of the key.
(POS. 8.10-11) Conversion factor for flow rate totalizers

[EUR/dm³+/- =X]

Setting of the conversion/currency value for direct (and reverse) flow totalizers. There are three
setting fields for this parameter, from left to right: 1) monetary symbol, 2) default/personalized
monetary symbol, 3) conversion coefficient. To modify, set the cursor over the required field. Setting
the monetary symbol can be achieved in two ways:

1.

Choosing one of the 7 predetermined monetary symbols (standard ISO 4217-REV81):
EUR=Euro; USD=USA dollar; CAD=Canadian dollar; AUD=Australian dollar; GPB=English
pound; CHF=Swissfranc; JPY=Japanese yen.

2.

Entering a three character (numbers or letters) personalized currency. To change the
characters, the cursor must be set on the symbol “/” (second field)

CURRENT VALUES IN mA ASSOCIATE TO THE % VALUE OF FULL SCALE

POSSIBLE FIELD

REVERSE FLOW VALUE

ZERO

DIRECT FLOW VALUE

d -110%

-100% 0%

+100%

t+110%

OutmA = 0 ÷ 20 +

0

0

0

20

20

OutmA = 0 ÷ 22 +

0

0

0

20

22

OutmA = 4 ÷ 20 +

4

4

4

20

20

* OutmA = 4 ÷ 22 +

4

4

4

20

22

OutmA = 0 ÷ 20 -

20

20

0

0

0

OutmA = 0 ÷ 22 -

22

20

0

0

0

OutmA = 4 ÷ 20 -

20

20

4

4

4

OutmA = 4 ÷ 22 -

22

20

4

4

4

OutmA = 0 ÷ 20

20

20

0

20

20

OutmA = 0 ÷ 22

22

20

0

20

22

OutmA = 4 ÷ 20

20

20

4

20

20

OutmA = 4 ÷ 22

22

20

4

20

22

OutmA = 0 ÷ 20 –0+

0

0

10

20

20

OutmA = 0 ÷ 22 –0+

0

1

11

21

22

** OutmA = 4 ÷ 20 –0+

4

4

12

20

20

OutmA = 4 ÷ 22 –0+

4

4.8

12.8

20.8

22

* Example 1: out 4-22 +

** Example 2: out 4-20 –0+

22 mA

20 mA

-110% F.S.

-100% F.S.

+100% F.S. +110% F.S.

zero

4 mA

I (mA)

I (mA)

20 mA

-110% F.S.

-100% F.S.

+100% F.S. +110% F.S.

zero

12 mA

4 mA