Procedure – Fluid Components International FLT Series Rack Mount User Manual

Page 14

Advertising
background image

Doc. No. 06EN003250 Rev. N/C

D - 3

FLT

Series FlexSwitch

Rack Mount

APPENDIX D - TEMP COMP

FLUID COMPONENTS INTL

Figure D-1. Potentiometer and Test Point Locations for Temp Comp

Procedure

1.

Turn off the instrument power.

2.

Install the instrument into the pipe or a test stand where it can be calibrated. Start the process media flowing at a
normal rate. Cool the process media to the lowest temperature in the expected operating range.

3.

Disconnect the wires leading to the sensing element (ACT, REF, COM HTR, HTR). Removal of the terminal
strip from the enclosure may be necessary for access to the wires.

4.

Connect the DMM's and the power supply to the sensing element as shown in Figure D-2.

5.

Set the power supply voltage to the proper voltage as shown in Table D-1. Turn on the power supply and check
the voltage setting.

Table D-1. Heater Voltage Settings

POWER SUPPLY SETTINGS

FLT93-S

3 Watts

1.75 Watts

0.75 Watts

0.27 Watts

Set For 18.0 Vdc

Set For 13.8 Vdc

Set For 9.0 Vdc

Set For 4.9 Vdc

FLT93-F

0.75 Watts

0.52 Watts

0.49 Watts

0.20 Watts

Set For 18.0 Vdc

Set For 17.0 Vdc

Set For 15.0 Vdc

Set For 10.6 Vdc