Fluid Components International FLT93 Nuclear User Manual

Page 51

Advertising
background image

FLT

®

Series FlexSwitch

TM

APPENDIX C - TEMP COMP

Fluid Components International LLC

51

POWER SUPPLY SETTINGS

FLT93-S

3 Watts

1.75 Watts

0.75 Watts

0.21 Watts

Set For 18.0 Vdc

Set For 13.8 Vdc

Set For 9.0 Vdc

Set For 4.9 Vdc

FLT93-F

0.57 Watts

0.52 Watts

0.40 Watts

0.25 Watts

Set For 18.0 Vdc

Set For 17.0 Vdc

Set For 15.0 Vdc

Set For 11.8 Vdc

Table D-1. Heater Voltage Settings

8. Start the process media fl owing at the desired switch point velocity and at the low temperature, let the instrument stabilize for fi fteen

minutes.

9. Record the resistance values of the active and reference RTD’s at the low temperature.

10. Raise the temperature of the process media to the maximum expected temperature. With the instrument power on, let the instrument

stabilize for fi fteen minutes. The difference between the low and the high temperature should not exceed 100°F.

11. Record the resistance values of the active and reference RTD’s for the high temperature.

12. Calculate the Temp Comp factor with the formula shown below.

TEMP
COMP
FACTOR

=

∆R Low Temperature - ∆R High Temperature

(R Reference High Temperature) - (R Reference Low Temperature)

The Temp Comp factor is not to exceed ±0.041.

13. If the Temp Comp factor is within tolerance, turn off the power to the instrument and stop the process media if needed. Disconnect the

DMM’s and the power supply from the instrument. Reconnect the sensing element wires to the control circuit socket and reinstall the
socket in the enclosure if it was previously removed. Do not pinch the wires between the socket and the enclosure.

14. Look up the resistance values to adjust potentiometers R5 and R8 in the Temp Comp Factor table (Table D-2). Follow the procedure in the

Restoring Temp Comp adjustment section using the values found in the table below.

15. If the calculated Temp Comp factor exceeds the allowable tolerance by a small amount (±0.01), using the maximum Temp Comp factor

may make the instrument perform satisfactorily. However, if the factor is out of tolerance by more than ±0.01 then it will be necessary to
repeat the calibration to verify the result. Continue with the adjustment procedure if the second result is within tolerance.

Figure D-1. Sensing Element Calibration Connections

Advertising