5.2.12 Compact (Integral)

Ex-d version

Installation of

add-on module

Remove the rear cover, by loosening the

safety tap allen screw (M3) and turn the rear

cover counter-clockwise.

Remove the electronics using the holes

provided.

The add-on module is fitted at the bottom of
the converter insert as shown. The label text

on the add-on module should face up as

shown.
The add-on module is fitted with the connector

facing out of the enclosure.

The add-on module can be pressed into

position.
Connect the flat cable connector to the

module and power up. The add-on module is

automatically initialized. The menu's are now
visable on the display/keypad of the

Transmitter.

Remove the flat cable from the plate.

Remove the plate from the module bay.