Fluid Components International CMF Series Manual Installation of Electronics User Manual
Page 5
In
st
allation of signal con.
Fluid Components Intl
Document 06EN003327 Rev. -
35
FlexCOR
™
Model CMF Series
5.2.3 Compact(Integral)
IP 67
Installation/
exchange of
SENSORPROM
®
memory unit
The SENSORPROM
®
unit is delivered mounted
in the terminal box of the sensor as shown.
To remove the SENSORPROM
®
unit, the
following procedure must be followed:
Remove the terminal box cover of the sensor,
or remove the transmitter, if it has been installed.
Remove the connection plate in the terminal
box by unscrewing the earth grounding screw
in the center as shown.
1
3
4
2
The SENSORPROM
®
unit is located in the
bottom of the terminal box and can be removed.
Note the orientation.
To re-assemble, mount the connection plate in
the terminal box and tighten the earth grounding
screw.
The label on the SENSORPROM
®
unit must
face the wall bracket.
Mount the transmitter on the terminal box.