Section 1 Description

1.1 General Description

The Hach Model 2100P ISO Portable Turbidimeter (Figure 1) measures turbidity
from 0.01 to 1000 FNU. The instrument operates on four AA alkaline batteries or
with an optional battery eliminator. Rechargeable nickel-cadmium cells may be
used, but cannot be recharged in the instrument. The carrying case holds teh
instrument, all standard accessories, and the optional battery eliminator.

Figure 1 2100P ISO Turbidimeter and Accessories

Note: Avoid prolonged exposure to ultraviolet light and sunlight.

Note: When taking a reading, place the instrument on a level, stationary surface. It should not be held in

the hand during measurement.

1.2 Accessories

Accessories supplied with the turbidimeter include six sample cells; one sealed vial
each of: <0.1-FNU, 20-FNU, 100-FNU, and 800-FNU StablCal® Stabilized
Formazin Standards; 4 AA alkaline batteries; 15 mL of silicone oil; oiling cloth;
carrying case; instrument manual; and quick reference card.

1.3 Principle of Operation

The Model 2100P ISO Portable Turbidimeter operates on the nephelometric
principle of turbidity measurement. This instrument meets the design criteria
specified by DIN ISO 7027.

The ratio optical system (Figure 2) includes an LED lamp, a 90° detector to monitor
scattered light, and a transmitted light detector. The microprocessor calculates the