Hach-Lange 2100P iso - Portable Turbidimeter Instrument User Manual

Page 23

Advertising
background image

Section 2, continued

Page 23

4650098 operation.fm

Calibration

7. The display will show “S1” (with the 1 flashing) and 20 FNU or the value of the

S1 standard for the previous calibration. If the value is incorrect, edit using the
arrow keys.

8. Fill a clean sample cell to the line with well mixed 20 FNU StablCal Standard or

20 FNU formazin standard. Insert the sample cell into the cell compartment by
aligning the orientation mark on the cell with the mark on the front of the cell
compartment. Close the lid.

9. Press

READ

. The instrument will count from 60 to 0 (67 to 0 if signal average is

on). The display will automatically increment to the next standard.

10. Remove the sample cell from the cell compartment.

11. The display will show “S2” (with the 2 flashing) and “100 FNU” or the value of

the S2 standard for the previous calibration. If the value is incorrect, edit using
the arrow keys.

12. Fill a clean sample cell to the line with well mixed 100 FNU StablCal Standard

or 100 FNU formazin standard. Insert the sample cell into the cell compartment
by aligning the orientation mark on the cell with the mark on the front of the cell
compartment. Close the lid.

13. Press

READ

. The instrument will count from 60 to 0 (67 to 0 if signal average is

on). The display will automatically increment to the next standard.

14. Remove the sample cell from the cell compartment.

15. The display will show “S3” (with the 3 flashing) and “800 FNU” or the value of

the S3 standard for the previous calibration. If the value is incorrect, edit using
the arrow keys.

16. Fill a clean sample cell to the line with well mixed 800 FNU StablCal Standard

or 800 FNU formazin standard. Insert the sample cell into the cell compartment
by aligning the orientation mark on the cell with the mark on the front of the cell
compartment. Close the lid.