Hach-Lange 2100P iso - Portable Turbidimeter Instrument User Manual

Page 22

Advertising
background image

Page 22
Calibration

4650098 operation.fm

Section 2, continued

To review a calibration, press

CAL

and then × to view the calibration standard

values. As long as

READ

is never pressed and

CAL

is not flashing, the calibration

will not be updated. Press

CAL

again to return to the measurement mode.

Note: For accuracy, use the same sample cell or four matched sample cells for all measurements during

calibration. Always insert the cell so the orientation mark placed on the cell during the matching
procedure is correctly aligned.

1. Rinse a clean sample cell several times with dilution water. Fill the cell to the

line (about 15 mL) with dilution water or use StablCal <0.1 FNU standard.

Note: The same dilution water used for preparing the standards must be used in this step.

2. Insert the sample cell in the cell compartment by aligning the orientation mark

on the cell with the mark on the front of the cell compartment. Close the lid.
Press

I/O

.

3. Choose signal average mode option (on or off) before pressing

CAL

. The

SIGNAL

AVERAGE

key is not functional in calibration mode.

4. Press

CAL

. The

CAL

and S0 icons will appear (the 0 will flash). The 4-digit display

will show the value of the S0 standard for the previous calibration. If the blank
value was forced to 0.0, the display will be blank. Press the right key to get a
numerical display.

5. Press READ. The meter will count from 60 to 0, (67 to 0 if signal average is on),

read the blank, and use it to calculate a correction factor for the 20 FNU standard
measurement. If the dilution water is >0.5 FNU, E1 will appear when the
calibration is calculated (See Preparing Dilution Water on page 21 for more
dilution water information). The display will automatically increment to the next
standard.

6. Remove the sample cell from the cell compartment.

The turbidity of the dilution water can be “forced” to zero by pressing the Ö key
rather than reading the dilution water. The display will show “S0 FNU”. Press
the × key to continue with the next standard.