Calibration

4650098 operation.fm

Section 2, continued

17. Press

READ

. The instrument will count from 60 to 0 (67 to 0 if signal average is

on). The display will increment back to the S0 display. Remove the sample cell
from the cell compartment.

18. Press

CAL

to accept the calibration. The instrument will return to measurement

mode automatically.

Note: If calibration errors occurred during calibration, error messages will appear after CAL is pressed.

Note: The standards must have a difference of at least 60 FNU.