Touring the Software

Click Apply to All Analytes to apply your selection to all selected analytes. Click OK to
change the default analysis to the analysis you selected. Click Cancel to close the dialog
box without saving. For more information, see “Analysis Settings” on page 25.

Units. Type the desired units for the analytes in this box.

Count. Type the desired bead count for the analytes by clicking in the Count box. If each
selected bead set does not acquire this number of events, a warning is added to the log
that not enough bead events were acquired. If you select bead sets that are not present,
the analyzer continues to acquire, trying to reach the number of events per bead for bead
sets that are not in the sample. Therefore, choose only the bead sets present in your
sample.

Apply All. Applies the information in the Units and Count boxes to all analytes.

Total Count. Select Stop after bead count reaches: to stop acquisition when the bead
count reaches a certain number determined by the user. Type the desired value in the box.
The default value is 100.

You can also specify the minimum allowable bead count per well that the xPONENT

®

software analyzes. This excludes data from any beads carried over during acquisition. For
information about this feature, see “Batch Options Tab” on page 79.

Selected Analytes List. Selected analytes appear in a list on the right side of the analyte
grid. This list includes the following information:

Name - The name of the analyte. Click and type to rename the analyte.

Analysis - To change the type of analysis for an analyte, click this field to open the
Analysis Settings dialog box and select another analysis from the list.