M2 Series

Instruction Manual 9R68-A

Mar, 2008

www.meriam.com

page 11 of 36

Units Select

The standard engineering units available on the M202 Precision
Absolute Manometer are:

Inches of Mercury at 0° C (in Hg @ 0° C)
Millimeters of Mercury at 0° C (mm Hg @ 0° C)
PSI
kPa
mbars
Bars
Torr
Feet (Altitude displays)
Meters (Altitude displays)


To change the engineering units the manometer should be “ON”
and in Measure Mode. Then follow these steps:

Keystroke

Display

1. Press the PRGM key.

Top line reads “PROGRAM
MODE” and bottom line reads
“UNITS SELECT”.

2. Press the PRGM key.

Top line reads “UNITS
SELECT” and bottom line shows
current engineering units.

3. Press the

or

arrow

key until desired engineer-
ing unit is displayed.

Engineering units on bottom line
of display change.

4. Press the PRGM key to
select the engineering unit.

Top line reads “PROGRAM
MODE” and bottom line reads
“UNITS SELECT”.

5. Press the

arrow key.

Bottom line reads “EXIT”.

6. Press the PRGM key.

Display returns to Measure Mode
in new engineering unit.