Meriam M202 Precision Absolute Manometer User Manual

Page 16

Advertising
background image

M2 Series

Instruction Manual 9R68-A

Mar, 2008

www.meriam.com

page 16 of 36

7. To change the value press
the

key to toggle the

display to “YES”.

Top line reads “VALUE=:
00000000”. Bottom line reads
“CHANGE?: YES”.

8. Press the PRGM key.

Top line reads “SEA LEVEL
METERS”. Bottom line reads
“00000000”.

9. Press the

or

arrow

keys to set the correct value
in the first digit.

Top line reads “SEA LEVEL
METERS”. Bottom line reads
“70000000”.

10. When the value is correct
press the PRGM key.

Cursor moves over to the next
digit.

11. Repeat steps 9 and 10
above until the correct
elevation is entered.

Top line reads “SEA LEVEL
METERS”. Bottom line reads
“700.0000”.

12. Press the PRGM key to
continue moving the cursor
to the right. When the last
digit is entered the unit will
leave the sea level select
mode and return to the
Program Mode.

Top line reads “PROGRAM
MODE”. Bottom line reads
“UNITS SELECT”.

13. Press the key. The
display will return to the
Measure Mode. When the
Correction to Sea Level is
ENABLED, the display will
have the letter “S” at the
beginning of the 2

nd

line.

Display reads;
“S 803.1 TORR”