Meriam M202 Precision Absolute Manometer User Manual

Page 31

Advertising
background image

M2 Series

Instruction Manual 9R68-A

Mar, 2008

www.meriam.com

page 31 of 36

Specifications

Type, Range and Display Resolution:

Absolute Isolated (AI) Type:
17 psia (900 mmHg) – XXX.YY
38 psia (2000 mmHg) – XXX.YY

Accuracy:
M202-AI0017: ±0.02 % F.S. (F.S. = 900 mm Hg)
M202-AI0038: ±0.015 % F.S.* from 0-1000 mm Hg

±0.025% F.S.* from 1000-2000 mm Hg

*F.S. = 2000 mm Hg Absolute

Includes the combined effects of temperature, linearity,
repeatability, hysteresis and resolution.

Warm up time = 5 minutes.

Temperature:

Storage = -40°C to +60°C (-40°F to +140°F)
Operating = -5°C to +50°C (23°F to +122°F)

Media Compatibility:

AI: Absolute pressure sensors for use with gases and liquids
compatible with 316L SS

Pressure Limits:

AI units: 77 PSIA (4000 mm Hg Abs)

Connection:

1/8” female NPT, 316L SS. P1 is the pressure connection.
P2 is not accessible (factory plugged with metal disc)

User must use a wrench on the pressure manifold when
installing user’s 1/8” NPT fitting. Do not tighten the
fitting without using a wrench on the pressure
manifold. Failure to use a wrench on the manifold will
damage the plastic enclosure and void warranty No
torque should be applied to the manifold with respect
to plastic enclosure.