4 overall noise / unsteady baseline – Multichannel Systems MEA Manual User Manual

Page 60

Advertising
background image

MEA Manual

60

9.4 Overall Noise / Unsteady Baseline

The baseline is unstable, signals are jumping or drifting.

Possible causes:

?

Bath electrode is not connected to ground.

 Connect the internal or external bath electrode to one of the ground inputs of the amplifier.

?

AgCl bath electrode needs is not well-chlorided.

 Rechloride the electrode or use a new one.

?

50 Hz hum: 50 Hz is the frequency of mains power in Europe. If the shielding and grounding of the setup
is not sufficient, electrical signals are picked up from the environment.

 Use a proper shielding. For example, you can place aluminum foil over the amplifier that is connected to

any metal part of the MEA amplifier. You can also use special shielding equipment like a Faraday cage.

The following screen shot shows a recording of a 60MEA200/30 without bath electrode and without
shielding
. You see that the signals are so high that the amplifier gets saturated, and you see a very strong
50 Hz hum.

The next pictures show the same MEA with bath electrode (silver pellet), but without shielding.
The baseline is very unsteady and oscillates with a frequency of 50 Hz.

Advertising