60stimmea – Multichannel Systems MEA Manual User Manual

Page 83

Advertising
background image

Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany

Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11

[email protected]
www.multichannelsystems.com

© 2013 Multi Channel Systems MCS GmbH

Product information is subject to change
without notice.

60StimMEA

60StimMEA200/30-ITO
60StimMEA200/30-Ti

Technical Specifications 60 Electrode MEA with 16 additional Stimulation Electrodes

Temperature

compatibility

0

-

125

°C

Dimension (W x D x H)

49 mm x 49 mm x 1 mm

Base

material

Glass

Track

material

Indium

tin

oxide

(ITO)

or

Titanium

(Ti)

Contact

pads

Indium

tin

oxide

(ITO)

or

Titanium

nitride

(TiN)

Electrode

diameter 30

μm

Interelectrode distance (centre to centre)

200 μm

Electrode

height

Planar

Electrode

type

Titanium

nitride

(TiN)

electrodes

Isolation

type

Silicon

nitride

500

nm

(PEVCD)

Electrode

impedance

30

-

50

k



Electrode

layout

grid

8

x

8

and

additional

4

pairs

of

large

stimulation

electrodes

(STIM

=

70

x

250

μm)

4

pairs

of

small

stimulation

electrodes

(S

=

30

μm)

Number of recording electrodes

60

Number

of

reference

electrodes

Without

internal

reference

electrode

MC_Rack
Source layout in “Data Source Setup”

2 dim. (MEA)

Channel

map

Default

MEA

perfusion

chamber

(w/o)

without

ring

(gr)

Glass

ring:

ID

+/-19

mm,

OD

24

mm,

height

6

/

12

mm

60 electrodes Microelectrode Array
with 16 additional stimulation electrodes.

Electrode layout: Standard 8 x 8 grid

For use with MEA-STIM-ADPT:

Adapter for MEAs with 16 additional
stimulation electrodes and for
MEA1060 amplifiers.

21

31

41

51

61

71

12

13

14

15

16

17

22

23

24

25

26

27

28

32

33

34

35

36

37

38

42

43

44

45

46

47

48

52

53

54

55

56

57

58

62

63

64

65

66

67

68

72

73

74

75

76

77

78

82

83

84

85

86

87

33 21 32 31 44 43 41 42 52 51 53 54 61 62 71 63

22

12

23

13

34

24

14

15

25

35

16

26

17

27

36 28 37 38 45 46 48 47 57 58 56 55 68 67 78 66

77

87

76

86

72

82

73

83

64

74

84

85

75

65

S

S

S

S

S

S

S

S

STIM

STIM

STIM

STIM

STIM

STIM

STIM

STIM

2450

μ

m

250 μm

7,755 mm

4,1 mm

7,755 mm

27,8 mm

Contact pad for stimulation adapter
MEA-STIM-ADPT
Diameter of contact pad 2 mm
Distance between contact pads 4.1 mm
(Centre to centre)

2000 μm

Advertising