10 synchronizing events, 1 digital output signals (sync out) – Multichannel Systems STG4004+STG4008 Manual User Manual

Page 61

Advertising
background image

57

10 Synchronizing

Events

10.1 Digital Output Signals (Sync Out)

The output of the digital Sync Out outputs is a 3.3 V TTL signal. A logic state of 1 means 3.3 V,
and a logic state of 0 means 0 V. Please note that the digital output is about 15 μs faster than
the analog output. This small offset is generally sufficient to make sure that the Sync Out signal
precedes the stimulation, which is important for synchronizing events.

You can program the output signal of the Sync Out channels in the same way as the analog
output signals, but due to the nature of a digital signal, only values of 0 and 1 are allowed.
A Sync Out signal is generated each time when the STG is started.

The output signal can be used for triggering external devices, for example, the data
acquisition with the MC_Card of the ME-System and the MEA-System. Please note that if you
use it for triggering the MC_Card, you should use output signals that have a duration of at least
200 μs. Otherwise, it can happen that the MC_Card ignores trigger events.

All four Sync Out outputs are active when the STG is started by pressing the Start / Stop button,
or by using the software command Start on the STG menu.

If you are using the Trigger In inputs for starting the STG, each of the four Sync Out channels is
controlled by a separate Trigger In channel. Sync Out 1 is active when Trigger In 1 is triggered,
Sync Out 2 is active when Trigger In 2 is active, and so on. The assignment of Sync Out channels
to Trigger In inputs can be customized, please see Trigger Settings.

If you use the Repeat feature in the Trigger Settings dialog box, the corresponding Sync Out
signal is repeated as well as the selected analog output signals.

You can, for example, synchronize four separate displays or other virtual instruments in MC_Rack,
one for each trigger input. You can also define a bit pattern in the Trigger Detector and trigger
instruments on this bit pattern, for example to activate virtual instruments only when trigger 1
and 3 are active at the same time. Please see the MC_Rack help or manual for more information
on the Trigger Detector and triggering.

Note:

No

Sync Out signal is sent when the STG is

stopped

on a trigger event or by pressing the

Start / Stop

button, that is, if the trigger event occurs or the button is pressed when the STG

has still been active.

Timing of blanking signal

If you have a MEA amplifier with blanking circuit in use, you can apply a blanking signal
(TTL pulse) for transiently switching off the input stage of the amplifiers during the stimulus,
thus avoiding the stimulus artifact.

MCS recommends a minimum blanking pulse that starts with the stimulus and stops 100 μs after
the stimulus. The fact that the digital output of the stimulus generator from MCS is about 15 μs
faster
than the analog output ensures that the blanking signal is delivered before the stimulus.
It is very important that the blanking pulse is delivered shortly before the stimulus.

Advertising