Analog Output Signals

79

The graphs were generated by a standard oscilloscope. The devices were set up according to
the recommendations under "Voltage Mode" and "Current Mode". The same pulse protocol
and the same channel number of the STG outputs were used for measuring the voltage and
the current output.

Example 1a: Voltage Mode: The green trace shows the voltage (+U) output; and the pink
trace shows the TTL output from the Sync Out that was used for triggering the oscilloscope

Rise time with programmed 500 mV and 50 μA pulses, 10 k

 load resistance.

Rise time (10 % / 90 %) = 3 μs