5. Align the drop tip with the opening in the Drop Counter. Put a graduated 10 mL cyl-

inder under the Drop Counter opening.

6. While on the Home screen, open the bottom stopcock to allow titrant to flow.

7. After about 10 mL of titrant is collected in the graduated cylinder, close the bottom

stopcock.

8. Record the exact volume from the graduated cylinder.

9. Build the desired display for you experiment.

10. Click the Experiment Tools button.

The Experiment Tools screen opens.

11. Click Calibrate Sensor.

The Calibrate Sensor: Select Measurement screen opens.

12. the Sensor: box and select Drop Counter.

13. Click the Measurement: box and select Fluid Volume (ml).

14. Click the Calibration Type: box and select 1 point (Adjust Slope Only).

15. Click NEXT.

The Calibrate Sensor screen opens.

16. Under Calibration Point 2 click the Standard Value: box and enter the recorded

volume (from step 8).

17. Under Calibration Point 2 click Read From Sensor.

18. Click OK to exit the Calibrate Sensor screen.

19. Click OK to exit the Experiment Tools screen.

Using sensor adapters

n

"Connecting a sensor through a Digital Adapter or Photogate Port": 19

n

"Connecting a sensor through an Analog Adapter": 20

Connecting a sensor through a Digital Adapter or Photogate Port

The Digital Adapter (PASCO part PS-2159) allows digital switch-type sensors such as
photogates and smart pulleys to be connected to a PASPORT interface and used with
SPARKvue. It also allows the use of a ScienceWorkshop Motion Sensor (CI-6742A) or
Rotary Motion Sensor (CI-6538).

Document Number: 012-11074D

19

SPARKvue

®

3   Setting up an experiment

Using sensor adapters