4

8510

VNA

system

Source

8360

Test set

851X

Receiver/detector

8510

processor/display

network

analyzer

Signal

separation

Stimulus

=

+

+

Figure 3. 8510 measurement systems are made up of modular components, the 8510
network analyzer, a test set and a compatible source.

The Agilent Technologies 8510 network analyzer consists of a family of
compatible products where each part is a separate system component. Each
complete system includes the 8510 network analyzer, a test set, and compatible
sources (measurement accessories are also needed to complete the measure-
ment setup), shown in Figure 3. The frequency range of the measurement sys-
tem is determined by the test set or the compatible source’s lesser frequency.
Example: if a 50 GHz test set is used with a 20 GHz source, the maximum
frequency measured is 20 GHz.