Mini-GASS User Manual | Original Instructions

PERMA PURE LLC

Mini-GASS™ User’s Manual
Doc. #212: Revision: 00A

Page 11 of 32

the equipment, within easy reach of the operator and shall be clearly marked as the
disconnecting device for the Mini-GASS.


Connect power supply line to terminal block located on the backplate of the control

enclosure inside the ½” conduit hub. (Refer to figure 5 for color coding). Connect the
system to an appropriate earth ground. Adhere to all electrical code requirements in effect
at the installation site. Connect power such that the protective earth conductor is the last to

take the strain.

Figure 5 – Electrical Connections

Alarm Relays

The SIVS2 has a 6 pole terminal block that will allow for field wiring of alarm relays for
connection of external alarm equipment (refer to Figure 6). Relay contacts are rated at 250

VAC, 3A, resistive. See Appendix F for additional information.

Figure 6 – Alarm Relays

Start-up procedure

Setup Check

1. Check that electrical, sample purge drain and /or steam connections have been

made.

2. Turn on compressed or instrument air to system.

Blue – Connect neutral or Line L2 here

Green/yellow – Connect ground earth
here

Brown – Connect Line L1 here