Mini-GASS User Manual | Original Instructions

PERMA PURE LLC

Mini-GASS™ User’s Manual
Doc. #212: Revision: 00A

Page 9 of 32

Heatless air dryer (optional)

1. Connect oil-free compressed air line to port labeled “Purge Gas Inlet”


2. A ¼” FNPT bulkhead fitting on bottom of enclosure is the heatless dryer purge air

exhaust, see figure 3. Humid air can be vented to atmosphere or piped to a
remote location.


Purge air exhaust (optional)
To vent the purge gas exhaust to a remote location, connect a line to the ¼” FNPT fitting
labeled “Purge Exhaust Port”. This is suggested for gas samples that are toxic, flammable or

otherwise hazardous. In case of failure of seal integrity, the sample will be vented to the
remote location. For runs under 10 ft, ¼” OD tubing is acceptable. For runs 10-40 ft,

3/8

OD tubing is recommended.

TY PE A GC

O N

FUSE

5.0 A.

120 VAC

7 0 0

E

S

U

F

POWER

0

O FF

I

CA L 32 00

*

O N

1

S

E

2

F

U

S

U

F

E

G as Analysis Sam p ling System

M ini-G.A.S.S.

EXH AUST TEM PERATU RE

PURGE AIR

T M

Increase

Dec rease

PRESSURE ADJ.

Increase

Decrease

25

5

SUPPLY PRESSURE

FLOWMETER

0

30

p si

15

10

20

O FF

0. 8

0. 4

12 .8

6. 4

3. 2

1. 6

51 .2

25 .6

ON

128

8

32

16

64

2

1

4

O FF

TIMER SETTINGS

FILTER DRAIN

0. 1
0. 2

ON

TIM E DE LA Y
IN M INU TE S

S W ITCH E S IN

TIM E DE LA Y

FO R TO T AL

O N P O SI TIO N

0 .25

0.5

C O MB IN E

OFF

IN HO UR S

TIME D E LA Y

20

10

15

5

25

30

PURGE FLO WRATE

SAM PLE INLET

SEA LIN G FITTIN G FO R

1

1

2

" O .D . H E ATED LINE

H EATED BAC KPLATE

HEA TLESS D R YER OP TION

BELOW FR ON T PAN EL

D IGITAL ELEC TR ON IC

TEM PE RA TU RE C O N TR O LLER

PO W ER SW ITC H

PO W ER C O N N EC TION

1

2

" C O ND U IT H UB

110/220 VAC

FU SE

D RY SAM PLE OU TLET
316 S S -

1

4

" N PT FEM ALE

PUR G E AIR INLET

1

4

" N PT FEM ALE

H EATLES S D RYER
PUR G E AIR E XHAU ST

1

4

" N PT FEM ALE

PU R GE EXH AU ST POR T

1

4

" N PT FEM ALE

N AFIO N M EM BR ANE
D RYER
M OD EL PD -200T-24SA

PU R GE EXH AU ST
TEM PER ATU R E SEN SOR

M EM BR AN E D R YER
PU R GE AIR FLOW M ETER

FLOW M ETER SU PPLY
PR ESSU R E R EGU LATOR

FLOW M ETER SU PPLY
PR ESSU R E GAU GE

34

3
4

"

12

5
8

"

12

1
8

"

29

5
8

"

Figure 3 - Mini-GASS Sampling System

Filter drain (optional)
Connect the line from the eductor outlet to designated collection/exhaust basin containing

acid absorption media to prevent release of exhaust to surrounding. ¼” i.d. tubing can be