15

Mini-GASS User Manual | Original Instructions

PERMA PURE LLC

Mini-GASS™ User’s Manual
Doc. #212: Revision: 00A

Page 15 of 32

6. Place new element in groove in bottom piece. Be sure that element is seated correctly

and parallel to glass shell.

7. Carefully mate bottom assembly onto top piece. Slight twisting motion may be required

to allow shell to slip over o-ring seal.

8. Visually make sure element is seated correctly in top groove.
9. Replace bolt through hole in bottom piece and screw clockwise into top piece. Do not

over-tighten center bolt. It should be just tight enough so it does not vibrate loose.

Over-tightening will not help the filter to seal.

Figure 7 – Filter Element Replacement

Dryers

Under normal conditions, Perma Pure dryers require little maintenance and can last for
several years. However, if there is no pre-filter and the tubing becomes clogged or

saturated with water, the dryer may require cleaning or repair. When disassembling the
dryer, note that the end fittings on the PD-SERIES dryer can be easily rotated on the shell

tube. This rotation should be avoided to prevent twisting the membrane tubes inside the
shell.


Refer to Appendix B for PD dryer element replacement.

Ammonia Scrubber

Media Replacement: When deposits are visible on 75% of the scrubber, scrubbing media
needs to be replaced.


1. Unscrew thumbscrew on bottom of housing.

2. Swing yoke to one side.

Advertising