System fine tuning

Temperature adjustment

 

Steam heated systems


Steam control valve is factory set to 180° F and is normally not field adjustable.
This temperature is adequate for most applications in which sample gas water

content is not in excess of 50% water vapor by volume. Consult factory if
condensation in sample tubing is detected.


Electrically heated systems

1. Check that no condensate is present in sample line between filter and dryer.

2. If water droplets are visible, increase temperature of filter/dryer compartment by

5° C.

3. Allow about ½ hour for system to stabilize and then check again for condensate.

4. If necessary, continue to increase temperature in 5° C increments until

condensate is no longer visible. Do not exceed 100° C! Maximum system

temperature is 100° C. Operation above this temperature can cause damage to
Nafion dryer. It is essential that purge gas flows continuously and

setpoint temperature is not above 100° C to prevent damage to dryer.

Maintenance

Filters

If system is fitted with a pre-filter, it should be checked regularly to ensure that the element
is in good condition. If element appears to be dirty or begins to cause flow restriction in
system, it should be replaced.


Filter Element Replacement

(Refer to Figure 7)

1. Loosen bolt on bottom of filter.
2. Gently pull apart assembly and remove old element.

3. Place new element into grooves in top and bottom of housing.
4. When re-assembling, inspect for o-rings on top and bottom caps and on center bolt.
5. Install glass shell onto bottom piece.