Mini-GASS User Manual | Original Instructions

PERMA PURE LLC

Mini-GASS™ User’s Manual
Doc. #212: Revision: 00A

Page 23 of 32

2. Connect cal gas line to fitting labeled “Cal Gas Inlet”.

3. Connect 110VAC power to blow back timer Note: This also provides power to the

filter in the probe.

Blow Back Operation

The blow back option includes a timer circuit which will initiate the blow back cycle based

on a regular, adjustable time period. The time cycle control turned to the far left will cycle the
blow back every 15 minutes. At the far right side, the period of time between cycles is about

24 hours. The period of time between blow back cycles should be calculated to occur as
often as when the filter has trapped one gram of particulates but no less frequently than

once per day.

A jumper placed between “AUTO OFF “and “COM” will disarm the timer. A contact
closure between terminals labeled “MAN B-B” and “COM” will initiate a blow back cycle.
The blow back solenoid valve will be opened for 2 seconds when the blow back cycle is

started.

Maintenance

Changing the Filter in Blow Back/Probe Assembly

1. Unscrew black filter cap inside Mini-GASS system (use gloves if hot to touch).

2. Filter is now exposed. With a pair of pliers, pull out the old filter.

3. Inspect o-rings which are at each end of filter. Replace if they are charred or deformed.

4. Place new filter element using pliers. Ensure it is centered and in contact with o-ring.

5. Screw cap back on the filter body.