Perma pure llc – Perma Pure Mini-GASS User Manual

Page 16

Advertising
background image



16

Mini-GASS User Manual | Original Instructions

PERMA PURE LLC

Mini-GASS™ User’s Manual
Doc. #212: Revision: 00A

Page 16 of 32

3. Separate housing and bottom cap as an assembly from top cap.
4. Remove spring and top screen.

5. Remove old media and dispose of properly (rinse housing with soapy water to clean).
6. Fill housing with 50cc of Berl Saddles (tap housing to allow material to settle).

7. Pour 150cc of Scrubbing Media (tap housing to allow material to settle).
8. Replace stainless steel screen on top of media.
9. Replace spring on top of screen.

10. Clean o-rings on shell and inside top manifold (replace if necessary).
11. Place center tube into o-ring seal in top cap.

12. Push and twist to seal housing around o-ring.
13. Replace yoke and finger tighten thumbscrew (do not over tighten).


Fuse


Fuse Replacement


1. Turn cap marked “FUSE” counter-clockwise ¼ turn and the fuse holder will come out.
2. Remove the blown fuse and replace with one of equal amperage rating. The standard

fuse is a BUSS type AGC or equivalent.

3. After installing correct replacement fuse, re-install fuse holder by pressing inward and

twisting ¼ turn clockwise.

Advertising