SKC 224-44XR Universal Pump User Manual

Page 12

Advertising
background image

10

Activating the Regulator

Remove the tamper-resistant cover.
Start the pump using the on/off
switch (Figure 2, #1), and adjust the
fl ow rate using the fl ow adjustment
screw (Figure 2, #4) until the built-
in rotameter reads approximately
1.5 L/min.

Remove the cap screw covering the
regulator valve (Figure 2, #11) and
turn the exposed screw four to fi ve
turns counterclockwise.

Replace the cap screw. The pump is now set for low fl ow.

Setting or Verifying Flow Rate

Ensure pump has run for
fi ve minutes before proceed-
ing with calibration.

Before use, allow pump to

equilibrate after moving

it from one temperature

extreme to another.

Connect a single adjustable
low fl ow holder (Figure 3)
to the pump intake (Figure
2, #6) using 1/4-inch Tygon
tubing.

Insert an opened sorbent tube
(Figure 3, #3) into the rubber
sleeve (Figure 3, #2) of the
low fl ow holder with the
arrow on the tube pointing
toward the holder.

Connect a calibrator to the exposed
end of the sorbent tube.

For low fl ow, turn valve screw

counterclockwise.

2

3

continued on page 10

Air

fl ow

Calibration train with tube

in low fl ow holder

1 Flow adjust screw

2 Rubber sleeve

3 Sorbent tube

Figure 3

Single Adjustable

Low Flow Holder with

sample tube