Low flow applications (5 to 500 ml/min), Setup – SKC 224-44XR Universal Pump User Manual

Page 14

Advertising
background image

12

Charger and battery

pack connected

Low Flow Applications (5 to 500 ml/min)

Using Multiple-tube Adjustable Low Flow Holder

Setup

For a diagram of the pump, see Figure 2, page 9.

Install battery (see Installing the Battery Pack on
page 16). For optimum charge, ensure pump is
not running. Charge the battery by connecting the
charger plug to the sampler charging jack (Figure 2,
#15). Ensure that the battery is fully charged before
sampling.

Aft er charging the batt ery pack, it is good practice to run the
pump for approximately fi ve minutes before calibrating. This
ensures the batt ery is in more steady-state conditions and
improves the agreement in pre and post-sampling calibrations.

Do not charge or operate pump from charger in hazardous locations.

Use only SKC-approved charger designated for this model to ensure
reliable performance. Failure to do so voids any warranty.

Ensure proper orientation of charging cable before plugging it into

the charging jack. Improper orientation/contact will short-circuit the

battery and voids any warranty.

Short-circuiting the battery pack will render it immediately inoperative.

Failure to follow warnings and cautions voids any warranty.

The battery pack may be kept on the SKC-approved charger for an indefi nite

time.

Figure 4

Quad Adjustable Low Flow Holder

1 Anti-tamper cover

2 Flow adjust screws

3 Rubber sleeve

4 Sorbent sample tube

5 Protective cover

Top view of fl ow

adjust screws