SKC 224-44XR Universal Pump User Manual

Page 6

Advertising
background image

4

Table 1. Pump Run Time in Hours with NiCad Battery

Following are typical run times achieved when using a fully charged nickel-

cadmium (NiCad) battery pack. Data is sorted by type of sample media. All run

times are listed in hours. Results obtained using a new pump and new fully

charged battery. Pump performance may vary.

** Filter back pressure exceeded pump capability during testing.

Increases in back pressure during sampling due to buildup of sample on
the fi lter can decrease batt ery life.

Mixed Cellulose (MCE) fi lter, 0.8-μm pore size

Flow Rate (L/min)

Filter Diameter

37 mm

25 mm

2.0

24.1

16.3

2.5

21.4

14.5

3.0

19.1

11.0

3.5

17.8

10.7

4.0

15.4

**

4.5

14.6

**

Polyvinyl Chloride (PVC) fi lter, 5.0-μm pore size

Flow Rate (L/min)

Filter Diameter

37 mm

25 mm

2.0

31.6

21.7

2.5

27.7

24.0

3.0

27.0

18.6

3.5

22.8

16.4

4.0

19.4

16.2

4.5

19.0

14.6

Note:

Table 2. Pump Run Time in Hours with NiMH Battery

Following are typical run times achieved when using a fully charged nickel-metal

hydride (NiMH) battery pack. Data is sorted by type of sample media. All run times

are listed in hours. Results obtained using a new pump and new fully charged

battery. Pump performance may vary.

Increases in back pressure during sampling due to buildup of sample on
the fi lter can decrease batt ery life.

Mixed Cellulose (MCE) fi lter, 0.8-μm pore size

Flow Rate (L/min)

Filter Diameter

37 mm

25 mm

2.0

37

33

2.5

34

26

3.0

31

21

3.5

29

18

4.0

25

15

4.5

20

14

Polyvinyl Chloride (PVC) fi lter, 5.0-μm pore size

Flow Rate (L/min)

Filter Diameter

37 mm

25 mm

2.0

47

41

2.5

38

33

3.0

35

30

3.5

26

27

4.0

22

25

4.5

21

23

Note: