Deactivating the Regulator

To ensure the pump is set for high fl ow,
remove the cap screw (Figure 1, #11)
covering the regulator valve and turn
the exposed screw clockwise until it
stops. (Do not overtighten.)

Replace the cap screw. The pump is
now set for high fl ow.

Setting or Verifying Flow Rate

Ensure pump has run for fi ve minutes before proceeding with calibration.

Before use, allow pump

to equilibrate after

moving it from one

temperature extreme to

another.

Using 1/4-inch Tygon

®

tubing, connect the sampling
medium to the pump intake
(Figure 1, #6).

Connect a calibrator to
the intake of the sampling
medium.

Remove the tamper-resistant cover. Start the pump using the on/off
switch (Figure 1, #1), and set the fl ow rate using the fl ow adjustment
screw (Figure 1, #4).

When the fl ow rate is set, turn off the pump and disconnect the
calibrator.

Replace the sampling medium used for calibration with an unexposed
medium for sample collection.

Calibration train with fi lter cassette

2

3

Flow

adjustment

screw

On/Off

switch

Tubing

Calibrator

Sampling

medium

For high fl ow,

turn valve screw

clockwise.