Sampling methods – System Sensor Pipe Installation User Manual

Page 11

Advertising
background image

User Guide: Aspirating Smoke Detector Pipe Installation

11

Sampling Methods

There are two general types of sampling methods: standard

pipe network sampling and capillary tube sampling. From these

sampling methods, several design configurations can be used

to meet the needs of a particular site environment. Local codes

and standards along with site requirements will help determine

the best air sampling method.

Standard Pipe Network Sampling

The following guidelines are used for any pipe network design.

Also, some additional guidelines specific to different sampling

methods may apply.

1. Local codes and standards always take precedence over

any values specified in this document.

2. Recommended pipe network material is nominal ¾"

schedule 40 internal diameter (25 mm OD) CPVC, PVC,

ABS or UPVC pipe.

3. Sharp bends decrease airflow and performance.

4. All pipe designs must have an end cap.

5. Multiple shorter pipes provide better performance than

a single longer pipe.

6. Symmetrical (Balanced) designs in pipe length, port size,

and port distribution are preferred to optimize FAAST

system performance.

Figure 11: Capillary tube sampling.

Figure 10: Sampling methods.

1.

2.

3.

4.

7. To prevent sampling ports form being blocked by dust

and dirt, taper and place the ports on the bottom side of the

sampling pipes, not on the top of the pipes. This ensures

that any falling debris does not clog the sampling ports.

8. To minimize the effect of pressure differentials, the sampled

air must be returned to the protected space wherever

possible. This eliminates any pressure differentials that

might reduce the air flow in the pipe network.

Capillary Tube Sampling

Capillary tube sampling is a method of locating sampling

ports remote from the main sampling pipe. This is particularly

useful where the main sampling pipe cannot be routed through

the area requiring protection for either technical or aesthetic

reasons. Capillary tubes are also used to sample equipment

cabinets or enclosures within the protected area.

In the absence of other guidance, it is recommended that a

minimum of two capillary sampling ports are sited in a room.

PipeIQ will allow sampling ports and capillary tubes to be added

as part of the design parameters. Local codes and standards

differ on issues, such as the minimum distance that detection

ports can be positioned from walls and ceilings. It is important

that the specific local regulatory requirements are observed.

The following guidelines are recommended for capillary tube use.

1. Try to keep the length of capillaries the same.

2. Capillary tube length should not exceed 26 ft. (8 m)

3. When sampling equipment cabinets or other enclosures,

the sampling port is typically placed at or close to the top of

the interior of the enclosure.

Frequently Asked Question – Can I paint the capillary port?

It is not recommended to paint the capillary port, but in certain

circumstances where aesthetics is a key feature, painting may

be allowed pending the approval of the local authority having

jurisdiction. If approved, the capillary port should be painted

and dried before the port is drilled. A sample port identification

label will be required to comply with NFPA.

Capillary Tube

Sampling Pipe

Tee Adaptor

Reducing Connector

Sampling Port

Ceiling Tiles

Advertising