Probes

LeCroy provides a passive probe for each WaveAce oscilloscope channel.

Probe Compensation

Passive probes must be compensated to flatten overshoot. This is
accomplished by means of a trimmer on the probe body.

1. Attach the connector end of your probe to any channel.
2. Connect the probe end to the CAL output connector at the front of

the oscilloscope.

3. Adjust the trim pot on the probe body until the square wave is as

flat as possible.

4. Set the Probe option attenuation to 10X by pressing the Channel

button and then the Probe menu button.

5. Set the switch to 10X on the probe.
6. Attach the probe tip to the PROBE COMP3V connector and the

reference lead to the PROBE COMP Ground connector. Press the
corresponding channel button, and then push the AUTO button.

7. Check the shape of the displayed waveform as follows:

Over-

Compensated

Compensated

Correctly

Under-

Compensated

Setting Probe Attenuation

Probes are available with various attenuation factors which affect the
vertical scale of the signal. Push the Channel button and then the Probe
menu option button.