Getting Started Manual

WA-GSM-E RevB

90

Press the SINGLE front panel button to begin loading the first .ads
firmware update file from your USB memory device Using the SAVE
ALL Screen
(on page 75). Otherwise, press the RUN/STOP front
panel button to exit the Update Firmware screen.

Locate the firmware upgrade files (the .ads files). Select the file
ending in _N.ads (this is the smaller of the two .ads files) and press
the Load button. The following status screen is shown.

The oscilloscope then beeps aloud twice and shows a screen stating
- The firmware has been updated completely, please boot up the
DSO again
. At this point, turn your oscilloscope power Off, wait a
short while, and then power it On.

When the oscilloscope has rebooted, return to the Utility → Update
Firmware
, press the SINGLE front panel button to begin loading the
second .ads firmware update file from your USB memory device
Using the SAVE ALL Screen (on page 75).

This time, select the file ending in _W.ads (this is the larger of the
two .ads files), and press the Load button.

Oscilloscope status screens and behavior is the same as the loading
of the _N.ads file including the need to power Off, wait a short
while, and then power it On your instrument completes the update.