Operation with a teledyne lecroy oscilloscope, Connecting the probe to the test circuit – Teledyne LeCroy ZS2500 User Manual

Page 16

Advertising
background image

ZS Series High-Impedance, Active Probes

Operation with a Teledyne LeCroy Oscilloscope

When the ZS1000, ZS1500, or ZS2500 probe is connected to any compatible

Teledyne LeCroy oscilloscope, the displayed scale factor and measurement

values are automatically adjusted.

The probe can be controlled through the oscilloscope graphical user interface.

When connected, a Probe dialog appears behind the corresponding channel

dialog. Refer to your oscilloscope’s manual for specific operation instructions.
Turning the Volts/Div knob controls the oscilloscope’s scale factor to give full

available dynamic range up to 2 V/div (16 V peak to peak). Turning the channel

Offset knob controls the probe input offset circuit over its range of ±12 V.

Connecting the Probe to the Test Circuit

To maintain the high performance capability of the probe in measurement

applications, care must be exercised in connecting the probe to the test circuit.

Increasing the parasitic capacitance or inductance in the input paths may

introduce a “ring” or slow the rise time of fast signals. Input leads which form a

large loop area will pick up any radiated electromagnetic field which passes

through the loop and may induce noise into the probe input.
Using one of the available accessories makes the ZS2500 probe with its small

profile and low mass head ideally suited for applications in dense circuitry.

10

924282-00 Rev A

Advertising
This manual is related to the following products: