Operator’s Manual

Verification Procedure

A. Output Zero Voltage

Output Zero Voltage Test Setup

1. Connect one end of a BNC cable to the female BNC connector on the probe

end of the ProBus extender cable. Connect the precision 50

Ω terminator to

the other end of the BNC cable.

2. Connect the banana plugs of the Precision terminator to the input of the

DMM. Make sure that the plug corresponding to the BNC shield (marked

"Ground") is connected to the LOW or COMMON input of the DMM.

3. Set the OFFSET on the oscilloscope to zero.
4. Set the DMM to read DC Volt on the most sensitive range.
5. Record the voltage measured on the DMM to 10 μV resolution as "Output

Zero Voltage" in the Test record.

6. Check that the voltage indicated by the DMM is between ±200

μV.

7. Disconnect the DMM from the precision 50

Ω terminator. Leave the

remaining setup in place for the next step.

924282-00 Rev A