Replacement parts, Performance verification, Required test equipment – Teledyne LeCroy ZS2500 User Manual

Page 24

Advertising
background image

ZS Series High-Impedance, Active Probes

Replacement Parts

The probe connection accessories and other common parts can be ordered

through the North America Customer Care Centers. Refer to Standard

Accessories table after the Introduction of this manual.

Performance Verification

This procedure can be used to verify the warranted characteristics of the

ZS1000, ZS1500, and ZS2500 High Impedance Active Probe.
The recommended calibration interval for the model ZS1000, ZS1500, and

ZS2500 is one year. The complete performance verification procedure should be

performed as the first step of annual calibration. Test results can be recorded on

a photocopy of the Test Record provided in Appendix A at the end of the

manual.
Performance verification can be completed without removing the probe covers

or exposing the user to hazardous voltages. Adjustment should only be

attempted if a parameter measured in the Performance Verification Procedure is

outside the specification limits.
NOTE: Adjustment should only be performed by qualified personnel
This procedure tests the following specifications:

• Output Zero Voltage

• LF Attenuation Accuracy

Required Test Equipment

The Standard Accessories table after the Introduction of this manual lists the

test equipment and accessories (or their equivalents) that are required for

performance verification of the ZS1000, ZS1500, and ZS2500 probes.
This procedure has been developed to minimize the number of calibrated test

instruments required.
Only the parameters listed in boldface in the Minimum requirements column

must be calibrated to the accuracy indicated.