Preliminary procedure, Operator’s manual – Teledyne LeCroy ZS2500 User Manual
Page 25
Operator’s Manual
Because the input and output connector types may vary on different brands and
models of test instruments, additional adapters or cables may be required.
List of Required Equipment
Description
Minimum Requirement
Test Equipment Examples
Digital Oscilloscope
ProBus Interface; Windows-
based
Teledyne LeCroy
WaveRunner
WaveSurfer Xs
Digital Multimeter
(DMM) with test probe
leads
4.5 digit
DC: 0.1% Accuracy
AC: 0.1% Accuracy
Agilent Technologies 34401A
Fluke 8842A-09
Function Generator
Sine Wave output amplitude
adjustable to 14.14 Vp-p (5
Vrms) into 1 MΩ at 70 Hz
Agilent Technologies 33120A
Stanford Research DS340
Power Supply
0-12 V, settable to 10 mV
HP E3611A
BNC Coaxial Cable
(2 ea.)
Male to
Male, 50 Ω, 36"
Cable
Pomona 2249-C-36
Pomona 5697-36
BNC Tee Connector
Male to Dual Female
Pomona 3285
Calibration Fixture
ProBus Extender Cable
Teledyne LeCroy
PROBUS-CF01
Terminator, Precision,
BNC
50 Ω ± 0.05%
Teledyne LeCroy
TERM-CF01
Banana Plug Adapter
(2 ea.)
Female BNC to Dual Banana
Plug
Pomona 1269
BNC to Mini-grabber
BNC Mail to Mini-grabber
Cable, 36”
Pomona 5187-C-36
Preliminary Procedure
1. Connect the ZS1000, ZS1500, or ZS2500 probe to the female end of
the ProBus Extension Cable. Connect the male end of the ProBus
Extension Cable to channel 1 of the oscilloscope.
2. Turn the oscilloscope on and allow at least 30 minutes warm-up time
for the ZS1000, ZS1500, or ZS2500 and test equipment before
performing the Verification Procedure.
924282-00 Rev A
19