<3. Component Names>

3-1

IM 01C27C01-01EN

3. Component Names

Cover flange

Process connector

Process connection

(Low pressure side)

Bolt

F0301.ai

Zero-adjustment screw

Ground terminal

Antenna (Note 3)

RF assembly

CPU assembly

Transmitter section

Amplifier Cover

Mounting

screw

Integral indicator

YES

(Write disabled)

NO

(Write enabled)

Hardware write protection switch (WR)

Write protection

Write protection

Switch Position

(Note 1)

(Note 2)

H

L

E

D

H

L

E

D

E WR

D

Not in use

Write protection switch

Slide

switch

Note 1: Set the switch as shown in the figure above to set the write protection. The hardware write protection switch is set to E side. Set

to H side for the switch of not-in-use.

Note 2: When the switch is D side (write protection setting), provisioning is acceptable. For details of provisioning, refer to section 7.4

“ Connecting to the Field Wireless Network “.

Note 3: The detachable antenna is applied when the amplifier housing code 7 or 8 is specified.

Figure 3.1

Component Names

Table 3.1

Display Symbol

Display Symbol

Meaning of Display Symbol

The output signal being zero-adjusted is increasing.

The output signal being zero-adjusted is decreasing.
Write protect function is enabled.