6 mounting the flushing connection ring, 1 mounting to pressure detector section, 2 mounting to process flange – Yokogawa EJX210B User Manual

Page 21: Mounting the flushing connection ring -3 4.6.1, Mounting to pressure detector section -3, Mounting to process flange -3, Important

Advertising
background image

<4. Installation>

4-3

IM 01C27C01-01EN

4.6 Mounting the Flushing

Connection Ring

4.6.1 Mounting to Pressure Detector

Section

The flushing connection ring is mounted to high

pressure side pressure detector section as shown

in Figure 4.5.
At the factory shipment, the flushing connection ring

is already assembled and attached to high pressure

side process detector section.

F0405.ai

Pressure-detector section

Drain/vent plug

Ring

Ring holder

Spiral gasket

View from pressure detector section side

Groove for installing

spiral gasket

Figure 4.5

Mounting to Pressure Detector Section

(1) Mount the ring holder on the ring and loosely

tighten the mounting screws.

(2) Place the spiral gasket in the ring groove. With

the ring correctly aligned and flush with the face

of the pressure detector, securely tighten each

ring holder’s mounting screws.

(3) Position the ring so that the drain/vent plugs are

aligned straight up and down.

4.6.2 Mounting to Process Flange

Tighten the bolts to completely close the gap

between the ring and the pressure detector section.
The mating flange, gasket, stud bolts and nuts are

to be procured by the customer.

F0406.ai

Ring

Diaphragm

Pressure-detector

section

Gasket

Mating flange

Spiral gasket

Figure 4.6

Mounting to Process Flange

IMPORTANT

• Confirm that there is no gap between the ring

and the process-detector section after they

are mounted on the process flange. A gap

can lead to a sudden, explosive release of

process fluids.

• When mounting or removing the ring,

take care not to tilt the pressure detector

downward as the ring can slip off and cause

injury.

• When re-mounting the ring, use the new

spiral gasket as shown in below table.

Table 4.1

Spiral Gasket for Pressure Detector

Section Side*

Part number

Size

Description

F9350SV

ш100Чш120Чt4.5 For 3-inch flange

F9970XF

ш100Чш120Чt4.5 For 3-inch flange**

F9350ST

ш70Чш90Чt4.5

For 2-inch flange

F9970XD

ш70Чш90Чt4.5

For 2-inch flange**

F9346ZH

ш60Чш75Чt4.5

For 1 1/2-inch flange

F9970XB

ш60Чш75Чt4.5

For 1 1/2-inch flange**

*:

Material; 316SST (Hoop), PTFE Teflon (Filler)

**:

For oil-prohibited use (Option code: /K1, /K5)

Advertising